Animal Husbandry and Feed Science ›› 2013, Vol. 34 ›› Issue (6): 7-7.doi: 10.12160/j.issn.1672-5190.2013.06.004

Previous Articles     Next Articles

Study on the Determination Method of K Content in Potato by Using TAS-986 Atomic Absorption Spectrophotometer

LING Shu-li1 ,JIA Zhi-qiang2 (1.Institute of Animal Nutrition and Feed, Inner Mongolia Academy of Agricultural & Animal Husbandry Sciences, Hohhot 010031 ,China;2.Inner Mongolia Office of Beijing Persee Company, Hohhot 010030,China)   

  • Online:2013-06-20 Published:2013-06-20

Abstract: In order to find out a fast, simple and reliable detecting method, the content of K in potato was detected by using flame method and TAS-986 atomic absorption spectrophotometer. Results showed that the correlation coefficient (R2) of standard curve was 0.990 0, and the average recovery was 91.58%, and the precision and repeatability (RSD) was less than 10%. Therefore, the determination method of K content in potato had some advantages of reliability, high precision, good repeatability and high recovery.

CLC Number: